GM1365-Temperature and Humidity Data Logger
  
1. Overview

  

Out of clusters of humidity and temperature instruments, this low power consumption Data-Logger combines the functions of signal-sampling, display, storage and analysis, widely be applied in the industries of medics, food, agriculture, laboratory, electronics, air-conditioning, weather service, archives management, archaeology study, fibre manufacturing and cold chain logistics, because this device can provide just-in-time solution of environm ent-monitoring, alerting and recording.

2. Main Features
  In Measurement : With high sensitivity detector to pick up temperature and humidity with quick response and high precison this device can review the readings of the current max,min ,up limit,low limit.
  In Memory : Automatically record and store the data of temperature ,humidity.
  Data Transfer : Directly connect the device with PC USB interface to transfer data with PC.
  In Programming : To set up sampling period,storage interval and alarming.
  In Analyzing : With specialized software to analyze the data of temperature and humidity with easy and reliable operation.
  In Safety : The bracket allows screw fastening with transparent water dust proof cover to protect the device from moisture and a up low limit can be set to alarm is overflow readings.
3. Technical Specifications
  

 
Temperature
Humidity
Measuring Range
-30-80 °C
0-100% RH
Memory Size
32256
Power
1/2AA 3.6V Lithium Battery
Size
126*28*22mm
Weight
49g
Accessories: Cover Braclcet,Screw 2,CD,Lithium Battery

 
Parameter
Condition
Min
Typical
Max
Units

Accuracy

typical
±0.3
°C
Tolerance
max
See Fig 2-1
See Fig 2-1
See Fig 2-1
°C
Repeatability
±0.1
°C
Response Time
t63%
5
30
s
Operating Range
extended
-40
125
°C
Long Term Drift
<0.04
°C/yr

                                         Table 2.1

Parameter
Condition
Min
Typical
Max
Units

Accuracy

Tolerance

typical
±0.2
% RH
max
                       See Fig 2.2
% RH
Repeatability
±0.1
% RH
Hysteresis
±1
% RH
Non-Linearity
 <0.1
100
% RH
Response Time
t63%
8
  S
Operating Range
extended
0
% RH
Long Term Drift
normal
  <0.5
% RH/Yr

                                       Table 2.2

 

More related Pictures:

 

 


For more information about the system, please contact sales@sensor.in